The CL3112D is a high-precision, multifunctional, cost-effective three-phase standard energy meter developed in compliance with the Recommendation R46 issued by the International Organization of Legal Metrology (OIML). It is designed for the testing and calibration of energy meters, power terminals, and other electrical measuring instruments. It can also be used for the traceability of Class 0.05 and lower-grade energy meter verification devices as well as other metrological equipment.


Function
- It can measure the distortion factor of voltage and current for each phase, as well as analyze the 2nd to 64th harmonics.
- It is capable of measuring all waveforms specified in Recommendation R46, including peak-top waves, flat-top waves, odd harmonics, subharmonics, high-order harmonics, and zero-crossing waves.
Technical Parameters
| Voltage Measurement | |
| Voltage Range | 30 V … 720 V |
| Voltage Accuracy | ± 0.02% (30 V ≤ U ≤ U max) |
| Current Measurement | |
| Current Range | 0.3 mA … 120 A |
| Current Accuracy | ± 0.02% (10 mA ≤ I ≤ 120 A), ± 0.05% (1 mA ≤ I ≤ 10 mA), ± 0.2% (0.3 mA ≤ I ≤ 1 mA) |
| Phase Measurement | |
| Phase Range | 0.00° … 359.99° |
| Phase Accuracy | ± 0.01 (I ≥ 10 mA, U ≥ 30 V) |
| Frequency Measurement | |
| Frequency Range | 40 Hz … 70 Hz |
| Frequency Accuracy | ± 0.01 Hz |
| Power Factor Measurement | |
| Power Factor Range | – 1.0 … 0 … + 1.0 |
| Power Factor Accuracy | ± 0.0005 |
| Power Measurement | |
| Active / Reactive / Apparent Power Accuracy | ± 0.02% (30 V ≤ U ≤ U max, 10 mA ≤ I ≤ I max, PF ≥ 0.5) |
| Electrical Energy Measurement | |
| Active / Reactive Energy Meter Accuracy | ± 0.02% (30 V ≤ U ≤ U max, 10 mA ≤ I ≤ I max, PF ≥ 0.5) |
| Pulse Inupt / Pulse Output | 0 … 50 KHz |
| Harmonic Power / Energy | Harmonic order 2nd – 20th: ± 0.2% of reading + 0.1% of range; Harmonic order 21st – 64th: ± 0.6% of reading + 0.15% of range |
| Other Indicators | |
| Operating Power Source | AC 220 V ± 10%, 50 Hz ± 5% |
| Power Consumption / Warm-up Time | < 120 VA / < 30 mins |
| Dimensions (mm) | (W × H × D): 485 × 133 × 500 |
| Weight | Approx. 19 kg |
Meter Constant Table
| CL3112D power constants | |||||
|---|---|---|---|---|---|
| 60 V | 120 V | 240 V | 480 V | 600 V | |
| 0.01 A | 1.8 x 109 | 1.8 x 109 | 1.8 x 109 | 1.8 x 109 | 1.8 x 109 |
| 0.02 A | 1.8 x 109 | 1.8 x 109 | 1.8 x 109 | 1.8 x 109 | 1.8 x 109 |
| 0.05 A | 1.8 x 109 | 1.8 x 109 | 1.8 x 109 | 1.8 x 109 | 1.8 x 109 |
| 0.1 A | 1.8 x 109 | 1.8 x 109 | 1.8 x 109 | 1.25 x 109 | 1.0 x 109 |
| 0.2 A | 1.8 x 109 | 1.8 x 109 | 1.25x 109 | 6.25 x 108 | 5.0 x 109 |
| 0.5 A | 1.8 x 109 | 1.0 x 109 | 5.0 x 108 | 2.5 x 108 | 2.0 x 108 |
| 1 A | 1.0 x 109 | 5.0 x 108 | 2.5 x 108 | 1.25 x 108 | 1.0 x 108 |
| 2 A | 5.0 x 108 | 2.5 x 108 | 1.25 x 108 | 6.25 x 107 | 5.0 x 107 |
| 5 A | 2.0 x 108 | 1.0 x 108 | 5.0 x 107 | 2.5 x 107 | 2.0 x 107 |
| 10 A | 1.0 x 108 | 5.0 x 107 | 2.5 x 107 | 1.25 x 107 | 1.0 x 107 |
| 20 A | 5.0 x 107 | 2.5 x 107 | 1.25 x 107 | 6.25 x 106 | 5.0 x 106 |
| 50 A | 2.0 x 107 | 1.0 x 107 | 5.0 x 106 | 2.5 x 106 | 2.0 x 106 |
| 100 A | 1.0 x 107 | 5.0 x 106 | 2.5 x 106 | 1.25 x 106 | 1.0 x 106 |
The output frequency for end of range in voltage/current pair with three phase connection is 50 kHz. The power pulses are on TTL/CMOS level. The burden capacity is > 20 mA.
Calculation formulas
Frequency

UR = voltage range
IR = current range
c = reference meter constant
Constant

The meter constants can be set in the settings menu or selected automatic. Setting range is 1 i/kWh to 2 x 109 i/kWh.