Single phase meters with closed I-P links
When do I need Isolation Current Transformers? require galvanic isolation for testing. This galvanic isolation is done by a Multi Secondary Voltage Transformer (MSVT). The MSVT has one secondary taping for each single phase meter under test.

The CLOU CL6202-xx can be configured for various secondary tappings. For requests, please indicate the no. of meter positions.
| item | value |
|---|---|
| No. of windings | max. 48 |
| Primary voltage | 220 V to 240 V |
| Primary frequency | 50 Hz or 60 Hz |
| Secondary voltage | 220 V to 240 V |
| Range of secondary burden | 4 VA to 15 VA |
| Accuracy Error prim / sec | ≤ ± 0,1% |
| Phase Error prim / sec | ≤ ± 2′ |
| Accuracy Error sec / sec (within burden range) | ≤ ± 0.01% |
| Phase Error sec / sec (within burden range) | ≤ ± 1′ |
Each secondary tapping is calibrated individually.