If a meter has no possibility to open the internal link between voltage and current, the test must be performed with an Isolated Current Transformer (ICT) at each measurement position.
The CL2030-D has a built-in electronic compensation. The ICTs can be remote controlled by PC. A safety protection by controller is built in. This assures that the secondary side is closed in case of high burden or open circuit. Each ICT will be delivered with a company calibration certificate. A certificate issued by an ILAC accredited laboratory is optional available.
Specifications ICT CL2030-D
Item | Value |
---|---|
Nominal Current | 100 A, max. load 120% |
Ratio | 1 : 1 |
Ratio error | ± 0.01% (0.2 A to 120 A) ± 0.03% (0.05 A < 0.2 A) ± 0.05% (0.01 A < 0.05 A) |
Phase displacement | ± 0.3'(0.2 A to 120 A) ± 3'(0.05 A < 0.2 A) ± 8'(0.01 A < 0.05 A) |
Max. rated secondary voltage | 0.7 V |
Max. rated burden for 10 mA ≤ I < 2 A | 200 mΩ |
Max. rated burden for I ≥ 2 A | 0.7 /I |
Frequency range | 45 Hz to 65 Hz |
Weight: | 15 kg |
Accessories
The ICTs can operate stand-alone or together with a remote control by PC. For remote control with 3rd party equipment, a control box CL2030-3D-CB is needed. An interface-description is available on request.
The links are uses to interconnect the primary side of ICTs. The copper links are galvanized. These links are designed to connect ICTs with a centre to centre distance of min 297 mm and max 317 mm.
If you are test equipment manufacturer you can get the ICT2030-D with your company-logo and without any CLOU indication from outside visible.